COMMAND
KEYWORD
FIELD(s)
FIELD
FORMAT
FIELD
DATA
RANGE
DESCRIPTION
a
NR1
0 to 4
Sets that signals in VPA
v
may be adjusted for the voltage drop across the A
terminals, or the current in the V terminals
a
=0: No adjustments are made
a
=1: adjusts the A signal to compensate for the V input current when V is
connected at the source and you require the results at the source (the V
current is added to the A signal)
a
=2:
adjusts the V signal to compensate for the A input drop when V is
connected at the load and you require the results at the source (the A drop
is added to the V signal)
a
=3:
adjusts the V signal to compensate for the A input drop when V is
connected at the source and you require the results at the load (the A drop
is subtracted from the V signal)
a
=4:
adjusts the A signal to compensate for the V input current when V is
connected at the load and you require the results at the load (the V current
is subtracted from the A signal)
BANDWIDTH
v
VDEF
‐
Sets the Bandwidth Limit for VPA
v
h
=0: Auto‐Track Bandwidth Limiting
h
=1: No Bandwidth Limiting
h
=2: Fixed frequency bandwidth limiting set by the
freq
field in Hz
The
freq
field is optional and is ignored if present if
h
is not 2.
h
NR1
0 to 2
freq
NR3
>0.0
CHANNELS
v
VDEF
‐
Sets the channels included in VPA
v
to
c
= additive channel selection,
CH1=1, CH2=2, CH3=4, CH4=8
c
NR1
0 to 15
COUPLE
v
VDEF
‐
Sets the default measurement result coupling for VPA
v
to AC+DC (
c
=0), AC
(
c
=1) or DC (
c
=2)
c
NR1
0 to 2
DIGITS
v
VDEF
‐
Sets the number of displayed digits for VPA
v
results to
d
d
NR1
3 to 6
EFFGROUP
v
VDEF
‐
Sets the efficiency group for VPA
v
to None (
e
=0), IN (
e
=1), MIDDLE (
e
=2)
or OUT (
e
=3)
e
NR1
0 to 3
FUND
v
VDEF
‐
Sets the fundamental frequency for VPA
v
f
=0: No fundamental (
freq
ignored)
f
=1: Fixed fundamental (
freq
sets the frequency)
f
=2: V input measurement (
freq
sets the maximum allowed frequency)
f
=3: A input measurement (
freq
sets the maximum allowed frequency)
f
=4: MAINS fundamental (
freq
ignored)
f
=5: AVIONICS fundamental (
freq
ignored)
f
=6 through 8: Use VPA1 through 3 resp. as the fundamental frequency
(
freq
ignored) – cannot use VPA
v
f
=9 through 11: Use GFUND1 through 3 resp. as the fundamental frequency
(
freq
ignored) – only valid if option MU is installed
The
freq
field need not be present if not needed
f
NR1
0 to 11
freq
NR3
Set by PERIOD
command
selection
HARMS
v
VDEF
‐
Sets the maximum number of harmonics for VPA
v
to
h
h
NR1
0 to 100 (0 to 500
if H500 installed)
PERIOD
v
VDEF
‐
Sets the measurement period for VPA
v
p
=0: VLF
p
=1: LF
p
=2: 10Hz/0.3s
p
=3: 20Hz/0.1s
p
=4: 45Hz/20ms
p
=5: 150Hz/10ms
p
=6: 500Hz/2ms (not valid unless the VPA contains W type channels)
p
=7: Full synchronization with VPA1 (not valid if
v
= VPA1)
p
=8: Full synchronization with VPA2 (not valid if
v
= VPA1 or VPA2)
p
NR1
v
= 1 : 0 to 6
v
= 2 : 0 to 7
v
= 3 : 0 to 8
RESPONSE
v
VDEF
‐
Sets the measurement response for VPA
v
to Fastest (
r
=0), Medium (
r
=1),
Slow (
r
=2) or Slowest (
r
=3)
r
NR1
0 to 3
STBYA
v
VDEF
‐
Sets VPA
v
for EN50564 5.3.2 method A standby power measurements
d
is the minimum number of days for the measurement.
h
is the minimum number of hours for the measurement.
m
is the minimum number of minutes for the measurement.
s
is the minimum number of seconds for the measurement.
d
NR1
0 to 99
h
NR1
0 to 99
m
NR1
0 to 99
s
NR1
0 to 99
STBYD
v
VDEF
‐
Sets VPA
v
for EN50564 5.3.2 method D standby power measurements
s
is the starting power detection level.
e
is the ending power detection level.
s
NR3
≥0.0
e
NR3
≥0.0 and ≤
s
WIRING
v
VDEF
‐
Sets the wiring method for VPA
v
w
=0: N x 1ø method.
w
=1: 2ø3w method.
w
=2: 3ø3w (2 channel) method.
w
=3: 3ø3w (3 channel) method.
w
=4: 3ø4w method
w
NR1
0 to 4