SCPI Commands for Digital Interfaces
TM-XDOP-01XN
4–33
AC Off Protection
The user has the option of setting whether the AC Off protection mechanism is
latched or automatically resumes operation.
SENSE:VOLT:AC:PROT:LATCH {ON|1|OFF|0}
Where:
ON
or
1
means the supply will be latched in shut down if an AC Off condition
occurs, until the user re-enables the output.
and
OFF
or
0
means the supply will automatically resume operation when the AC line
input returns to normal.
The default value is OFF.
To query if AC Off has tripped:
SENSE:VOLT:AC:PROT:TRIP?
Clear Protection Event
To clear a protection mechanism that has tripped:
OUTP:PROT:CLE
This will clear all protection mechanisms and re-enable the output. If the
condition that caused the alarm still exists, the protection will be allowed to trip
again.
View Power Supply Output
The following 3 commands query the voltage, current, or power being supplied at
the output terminals.
To measure the voltage output:
MEAS:VOLT?
To measure the current output:
MEAS:CURR?
To measure the power output:
MEAS:POW?
TM-XDOP-01XN.book Page 33 Monday, July 17, 2006 11:19 AM
Summary of Contents for XDC 10-1200
Page 2: ......
Page 3: ...XDC 6000 Watt and 12000 Watt Series Digital Programmable Power Supply Operating Manual ...
Page 8: ...vi ...
Page 12: ...x ...
Page 22: ...xx ...
Page 24: ...xxii ...
Page 48: ...Installation 2 10 TM XDOP 01XN Figure 2 5 Attaching the AC Input Wires for 6000 Watt units ...
Page 114: ...Operation 3 52 TM XDOP 01XN ...
Page 208: ...A 16 ...
Page 224: ...B 16 ...
Page 259: ...D Error Messages Appendix D lists all the error codes and their meanings ...
Page 268: ...D 10 ...
Page 283: ......