Entering Calibration Mode
TM-XDOP-01XN
B–5
Security code
To protect calibration data, a security code is required to enter calibration mode.
The security code set at the factory to "
0000
."
The password can be changed from the remote interface or the front panel.
Calibration state must be ON to change the password. From the front panel, select
"
Change Code
" from the calibration menu and enter the new code.
To change the Calibration security code:
If you have entered a valid code, the prompt displays
Output V Cal
. You can
either continue working with the calibration options or press
EXIT
to leave
calibration mode.
The SCPI command to change the security code is:
CAL:CODE <security_code>
The security code is any 4-digit number enclosed by quotation marks. Trying to
change the password to an invalid one causes an error.
Step # Do This
You Will See
1
Output V Cal
2
or
Scroll to select
Change Code
.
Change Code
3
Enter a new 4-digit code.
Code ####
4
You are back in the Calibration menu.
Output V Cal
ENTER
TM-XDOP-01XN.book Page 5 Monday, July 17, 2006 11:19 AM
Summary of Contents for XDC 10-1200
Page 2: ......
Page 3: ...XDC 6000 Watt and 12000 Watt Series Digital Programmable Power Supply Operating Manual ...
Page 8: ...vi ...
Page 12: ...x ...
Page 22: ...xx ...
Page 24: ...xxii ...
Page 48: ...Installation 2 10 TM XDOP 01XN Figure 2 5 Attaching the AC Input Wires for 6000 Watt units ...
Page 114: ...Operation 3 52 TM XDOP 01XN ...
Page 208: ...A 16 ...
Page 224: ...B 16 ...
Page 259: ...D Error Messages Appendix D lists all the error codes and their meanings ...
Page 268: ...D 10 ...
Page 283: ......