Operation
3–6
TM-XDOP-01XN
4.
OUT ON/OFF:
This is a toggle key that enables and disables the power
supply output. Normally, you should leave the power supply in its Output Off
state when no load is attached or there is no need for DC output. (See “Turn
Output On or Off” on page 3–9.)
When the Out On annunciator is illuminated, the output is on. When the Out Off
annunciator is illuminated, the output is off.
5.
LCL/RMT:
This key lets you toggle between local and remote control. If the
RMT annunciator is illuminated, press
LCL/RMT
to return control to the
front panel, if local mode has been enabled by the controller. (See “Toggle
Local/Remote” on page 3–18.)
6.
PROT SET:
Lets you view and set protection setpoints. (See “Set Output
Protection” on page 3–10.)
7.
RECALL:
Lets you apply stored power supply settings. (See “Recall
Settings” on page 3–23.)
8.
EXIT:
Lets you cancel an operation or leave Calibration mode or Auto
Sequence mode.
Menu Navigation
Four keys allow you to access many functions available on the menu. These keys
are
MENU
,
ENTER
, and the
Up
and
Down
arrow keys.
Top Level Menu Items
To display the first menu item, press
MENU
. To display the other top level menu
items in the order listed below, press
MENU
or the
Down arrow
repeatedly. To
display the other items in reverse order, press the
Up arrow
repeatedly.
The top level menu items are:
1.
ERROR MSGS:
Lists up to 50 queued errors. (See page 3–26.)
2.
USER LINES:
Configures auxiliary lines A and B. (See page 3–27.)
3.
PON CONFIG:
Configures the power-on settings. (See page 3–29.)
4.
S/D RECOVERY:
Sets up shutdown recovery options for AC Off and Over-
Temperature Protection. (See page 3–15.)
5.
REMOTE SELECT:
Sets up the remote access option. (See page 3–19.)
6.
REMOTE CONFIG:
Configures the remote access option. (See page 3–20.)
7.
AUTO SEQ PGM:
Programs automatic sequences. (See page 3–33.)
8.
CURRENT SHARE:
Sets up master/slave relationships for multiple-supply
configurations. (See Chapter 5, “Current Sharing”.)
9.
POWER SETPT:
Sets up the power output. (See page 3–9.)
TM-XDOP-01XN.book Page 6 Monday, July 17, 2006 11:19 AM
Summary of Contents for XDC 10-1200
Page 2: ......
Page 3: ...XDC 6000 Watt and 12000 Watt Series Digital Programmable Power Supply Operating Manual ...
Page 8: ...vi ...
Page 12: ...x ...
Page 22: ...xx ...
Page 24: ...xxii ...
Page 48: ...Installation 2 10 TM XDOP 01XN Figure 2 5 Attaching the AC Input Wires for 6000 Watt units ...
Page 114: ...Operation 3 52 TM XDOP 01XN ...
Page 208: ...A 16 ...
Page 224: ...B 16 ...
Page 259: ...D Error Messages Appendix D lists all the error codes and their meanings ...
Page 268: ...D 10 ...
Page 283: ......