Power Supply Operation
TM-XDOP-01XN
3–17
The possible alarms are:
Shutdown vs Protection Alarm
If a protection setpoint is exceeded, the system does the following:
1. If
S/D if tripped? Y
has been selected, the unit shuts down.
If it is an OVP alarm, the unit shuts down.
2. If
S/D if tripped? N
has been selected, and it is not an OVP alarm, the
unit does not shut down but does create an alarm message.
3. If the unit is not shutting down, the system still sets the appropriate status bits
in the questionable status register, which can be queried remotely. See “Status
Registers” on page 4–43 for more information.
•
OVP Shutdown
•
UVP Shutdown
•
OCP Shutdown
•
UCP Shutdown
•
OPP Shutdown
•
UPP Shutdown
•
UVP Alarm
•
OCP Alarm
•
UCP Alarm
•
OPP Alarm
•
UPP Alarm
•
Fold Shutdown
•
Sense Shutdown
•
Hi Temp Alarm
•
OTP Shutdown
•
AC Off
•
Slave Shutdown Alarm (12000 Watt only)