Remote Operation
4–24
TM-XDOP-01XN
Power On Output State
You may also change the output state whether the output is enabled or disabled at
power on.
To change the power on at output state:
OUTP:PON:STAT [ON|OFF|1|0]
To check the setting:
OUTP:PON:STAT?
Reset
Resetting the unit puts certain features to a known state. These states are listed in
the table below.
To reset the unit:
*RST or SYST:RES
Table 4-8
Features Affected by Reset (*RST) Command
Feature
Reset State (*RST)
Voltage setpoint
0.0V
Current setpoint
0.0A
Power setpoint
103% of power rating
Low voltage setpoint limit
0.0V
High voltage setpoint limit
103% of voltage rating
Low current setpoint limit
0.0A
High current setpoint limit
103% of current rating
Low power setpoint
0% of power rating
High power setpoint
103% of power rating
Over voltage protection
Disabled (0.0V)
Under voltage protection
Disabled (0.0V) and not shutdown when tripped
Over current protection
Disabled (0.0A) and not shutdown when tripped
Under current protection
Disabled (0.0A) and not shutdown when tripped
Over power protection
Disabled (0.0W) and not shutdown when tripped
Under power protection
Disabled (0.0W) and not shutdown when tripped
Fold shutdown protection
None and delay 0.5s
AC off shutdown recovery
Autorecover
Over temperature shutdown
recovery
Latched
Triggered voltage setpoint
Disabled (Default)
Triggered current setpoint
Disabled (Default)
TM-XDOP-01XN.book Page 24 Monday, July 17, 2006 11:19 AM
Summary of Contents for XDC 10-1200
Page 2: ......
Page 3: ...XDC 6000 Watt and 12000 Watt Series Digital Programmable Power Supply Operating Manual ...
Page 8: ...vi ...
Page 12: ...x ...
Page 22: ...xx ...
Page 24: ...xxii ...
Page 48: ...Installation 2 10 TM XDOP 01XN Figure 2 5 Attaching the AC Input Wires for 6000 Watt units ...
Page 114: ...Operation 3 52 TM XDOP 01XN ...
Page 208: ...A 16 ...
Page 224: ...B 16 ...
Page 259: ...D Error Messages Appendix D lists all the error codes and their meanings ...
Page 268: ...D 10 ...
Page 283: ......