BOD
163
17
Wait for the end of the measurement.
18
Repeat from step 12 to step 18 for all blank value
solutions.
Subsequently, the
End measurement
window opens
for the sample dilutions.
19
Check the data on the display.
l
If the data is correct: Continue with step 23.
l
To change the data: Continue with step 20.
20
Swap between the selection fields with
[Next]
.
21
Select
Yes
or
No
in the
Remove
field with
82
.
l
If
Remove No
was selected, the end measurement
is carried out.
l
If
Remove Yes
was selected, the end
measurement is not carried out. The value
appears on the report marked by *. This value is
not used for calculating the BOD value.
22
Immerse the sensor in the sample dilution.
23
Confirm the data with
[Start]
.
l
If
Remove Yes
was selected, a security prompt
appears before the removal. After confirming with
Yes
the next sample dilution is offered (see
l
If
Remove No
was selected, the measurement of
the sample dilution begins.