pH
89
6.3
Measuring the pH value
Preparatory
activities
Perform the following preparatory activities when you want
to measure:
Note
Incorrect calibration of pH electrodes leads to incorrect
measured values. Calibrate regularly. The measuring
system has to be calibrated before measuring.
Temperature probe
You can measure with or without a temperature probe. If a
temperature probe is connected, it is indicated on the
display by
v
.
Note
If you use a pH electrode without temperature probe, you
can also use the temperature probe of a D. O. probe or a
conductivity measuring cell (see section 5.5.2 M
). If you use the temperature probe of a D.
O. probe or a conductivity measuring cell,
y
or
z
appears
on the display.
1
Connect the electrode to the measuring module.
2
Call up the
Display
mode with
m
.
3
Call up the
pH
measured value display with
46
.
4
Adjust the temperature of the solutions and measure
the current temperature if the measurement is made
without a temperature probe.
5
With manual temperature input:
Use
82
to call up the temperature input field and
adjust the temperature.
Use
[ESC]
to terminate the temperature input.
6
Calibrate or check the measuring system with the
electrode.
7
Switch to the measuring mode with
m
.
1
Perform the preparatory activities.
Summary of Contents for inoLab 740
Page 18: ...Safety 18...
Page 27: ...Commissioning 27...
Page 32: ...Commissioning 32...
Page 50: ...Basic principles of operation 50...
Page 92: ...pH 92...
Page 134: ...Ion concentration 134...
Page 148: ...Dissolved oxygen D O 148...
Page 174: ...BOD 174...
Page 232: ...Data output 232...
Page 236: ...Maintenance cleaning disposal 236...
Page 270: ...Lists 270...
Page 274: ...Lists 274...