BOD
157
Measurement of
sample dilutions
28
Check the data on the display.
l
If the data is correct: Continue with step 32.
l
To change the data: Continue with step 29.
29
Enter a name for the sample dilution with
82
, an
external keyboard or via the barcode reader.
30
Swap between the entry fields with
[Next]
.
31
Enter the number of the sample dilutions in the
Number
field with
82
or an external keyboard.
32
Confirm the settings with
[Start]
.
The
Start measurement Dilution 1
window appears.
33
Check the data on the display.
l
If the data is correct: Continue with step 40.
l
To change the data: Continue with step 34.
34
Enter a name with
82
, an external keyboard or via
the barcode reader.
35
Swap between the entry fields with
[Next]
.
36
Enter the volume of the sample in the
Vol. sample
field with
82
or an external keyboard.
37
Swap between the entry fields with
[Next]
.