![wtw inoLab 740 Operating Manual Download Page 75](http://html1.mh-extra.com/html/wtw/inolab-740/inolab-740_operating-manual_1000561075.webp)
pH
75
6.1.5
Resetting to default settings
You can reset (initialize) the settings for the pH
measurement to the default settings.
In a reset, the following functions for the pH measurement
are reset to the default settings:
pH reset
With the reset pH function, all the settings for pH
measurements are reset to the default values.
Measuring mode
pH
Asymmetry U
Asy
in mV
0 mV
Slope S (k) in mV
-59.16 mV
Calibration procedures
AutoCal TEC
Temperature, manual
25 °C
1
Switch to the
Menu
mode.
2
Use
46
to select the main
Cal
menu.
3
Select the
Reset pH
menu item with
82
.
4
Confirm the selected menu item with
g
.
The default settings for pH measurements have been
restored.
The
Menu
mode is terminated.