BOD
158
38
Enter the total volume of the sample dilution in the
Vol.Total
field with
82
or an external keyboard.
39
Immerse the sensor in the sample dilution.
40
Start the measurement with
[Start]
.
The sample dilution is measured.
41
Wait for the end of the measurement.
42
Repeat from step 33 to step 42 for all sample
dilutions.
Subsequently, the
Routine STORE
window opens.
43
Select
Save Routine YES
or
No
in the field with
82
.
If
Save Routine
YES
was selected, the routine is
saved under a routine name.
44
Confirm the settings with
[Ok]
.
l
If
Save Routine NO
was selected, the start
measurement is ended.
l
If
Save Routine
YES
was selected, an entry field
for the routine name appears.