System settings
69
5.8
Modules
The calibration data can only be assigned to a user name if
the calibration data were recorded with the current
combination of terminal and module.
As a result of the registration of the modules, the terminal
recognizes whether the calibration data were stored in the
module with this or another terminal.
The registration of a module to a terminal is made
automatically in the first calibration of a module at a terminal.
If a module is registered at a terminal, the user name with
which the calibration was performed appears on the display
during the prompt of a calibration protocol, and the module
appears in the
Registered modules
list. This list contains all
registered modules with the module type and serial number.
If a module is not registered at a terminal, a note together
with
unknown
as the user name appears on the display
during the prompt of a calibration protocol.
Note
The maximum number of registered modules is 16. Further
modules are not registered. Calibrations are nevertheless
possible.
Anonymous
appears as the user name.
Viewing, printing out and editing the list of
Registered
modules
is only possible during login with the user name
Admin
.
Summary of Contents for inoLab 740
Page 18: ...Safety 18...
Page 27: ...Commissioning 27...
Page 32: ...Commissioning 32...
Page 50: ...Basic principles of operation 50...
Page 92: ...pH 92...
Page 134: ...Ion concentration 134...
Page 148: ...Dissolved oxygen D O 148...
Page 174: ...BOD 174...
Page 232: ...Data output 232...
Page 236: ...Maintenance cleaning disposal 236...
Page 270: ...Lists 270...
Page 274: ...Lists 274...