pH
83
8
Thoroughly rinse the electrode with deionized water.
9
If necessary, set the temperature of the second buffer
solution with
82
10
Immerse the electrode and the temperature probe in
the second buffer solution.
11
Start the measurement with
g
.
The
h
display indicator flashes. The electrode
voltage (mV) appears on the display.
As soon as a stable value is reached,
h
disappears.
The slope value (mV/pH) and the asymmetry value
(mV) appear on the display. The probe symbol shows
the calibration evaluation after the two-point
calibration.
12
Use
g
to go on to the three-point calibration
(
ct3
appears on the display)
or terminate after the two-point calibration with
m
(the current calibration data are stored).