![wtw inoLab 740 Operating Manual Download Page 76](http://html1.mh-extra.com/html/wtw/inolab-740/inolab-740_operating-manual_1000561076.webp)
pH
76
6.2
Calibrating pH measurements
Why calibrate?
pH electrodes age. This changes the asymmetry and slope
of the pH electrode. This may cause the measuring system
to ignore the limits of a calibrated system. Calibration
determines the current values of the asymmetry and slope
and stores them in the measuring module.
Calibration must be performed at regular intervals.
Note
Always calibrate after replacing the pH electrode.
You can choose between two calibration procedures:
AutoCal
and
ConCal
:
AutoCal
is a fully automatic calibration using the datasets selected
for buffer solutions in the
pH buffer
menu.
ConCal
is the conventional two-point
calibration method and single-
point calibration is the simplified calibration method.
AutoRead
When calibrating with AutoCal, the AutoRead function is
automatically activated.
The current AutoRead measurement can be terminated at
any time (accepting the current value) by pressing
g
.
Calibration record
The calibration record contains the calibration data of the
current calibration. You can call up the calibration record by
outputting the data storage (section 11.3).
Note
You can automatically print a calibration record following
calibration. To do so, switch on the printer before calibrating
(print LED lights up green). After a valid calibration, the
record is printed.
Summary of Contents for inoLab 740
Page 18: ...Safety 18...
Page 27: ...Commissioning 27...
Page 32: ...Commissioning 32...
Page 50: ...Basic principles of operation 50...
Page 92: ...pH 92...
Page 134: ...Ion concentration 134...
Page 148: ...Dissolved oxygen D O 148...
Page 174: ...BOD 174...
Page 232: ...Data output 232...
Page 236: ...Maintenance cleaning disposal 236...
Page 270: ...Lists 270...
Page 274: ...Lists 274...