What to do if...
246
14.6
Conductivity system messages
Measuring modules for the inoLab Level 3:
Cond, pH/Cond and Multi.
Error message
OFL
Error message
E3
Cause
Remedy
– Measuring cell not
connected
– Connect measuring cell
– Cable broken
– Replace measuring cell
– The measured value is
outside the measuring
range
Cause
Remedy
– Measuring cell
contaminated
– Clean cell and replace it
if necessary
– Calibration solution not
suitable
– Check calibration
solutions