What to do if...
239
14.3
pH/voltage system messages
Measuring modules for the inoLab Level 3:
pH, pH/ION, pH/Cond, pH/Oxi and Multi.
Error message
OFL
Error message
E3
Cause
Remedy
pH/voltage, electrode:
– Not connected
– Connect electrode
– Air bubble in front of the
diaphragm
– Remove air bubble
– Air in the diaphragm
– Extract air or
moisten diaphragm
– Cable broken
– Replace electrode
– Gel electrolyte dried out
– Replace electrode
Cause
Remedy
pH/ORP electrode:
– Diaphragm contaminated
– Clean diaphragm
– Membrane contaminated
– Clean membrane
– Moisture in the plug
– Dry plug
– Electrolyte out of date
– Replenish electrolyte or
replace electrode
– Electrode worn out
– Replace electrode
– pH electrode: broken
– Replace electrode
– ORP electrode: short-
circuit
– Replace electrode