Ion concentration
97
11
Thoroughly rinse the electrode with deionized water.
12
Immerse the electrode in the blank value solution.
13
Select
Yes
with
[Next]
.
14
Confirm
Yes
with
g
.
15
Terminate the blank value determination with
[ESC]
.
The
-
symbol no longer flashes. The displayed
measured value is reduced by the blank value.
The blank value correction is active until the blank
value is switched off.