pH
84
Three-point
calibration
Four-point
calibration
13
Thoroughly rinse the electrode with deionized water.
14
If necessary, set the temperature of the third buffer
solution with
82
(see section 5.5.1).
15
Immerse the electrode and the temperature probe in
the third buffer solution.
16
Press the
g
key.
The
h
display indicator flashes.
The electrode voltage (mV) appears on the display.
As soon as a stable value is reached,
h
disappears.
The probe symbol shows the calibration evaluation
after the three-point calibration.
The value of the slope S (in mV/pH) and the value of
the asymmetry (in mV) appear on the display.
17
Use
g
to go on to the four-point calibration
(
ct4
appears on the display)
or terminate after the three-point calibration with
m
(the current calibration data are stored).
18
Thoroughly rinse the electrode with deionized water.
19
If necessary, set the temperature of the fourth buffer
solution with
82
(see section 5.5.1).
20
Immerse the electrode and the temperature probe in
the fourth buffer solution.
Summary of Contents for inoLab 740
Page 18: ...Safety 18...
Page 27: ...Commissioning 27...
Page 32: ...Commissioning 32...
Page 50: ...Basic principles of operation 50...
Page 92: ...pH 92...
Page 134: ...Ion concentration 134...
Page 148: ...Dissolved oxygen D O 148...
Page 174: ...BOD 174...
Page 232: ...Data output 232...
Page 236: ...Maintenance cleaning disposal 236...
Page 270: ...Lists 270...
Page 274: ...Lists 274...