Conductivity
187
Accepting the cell
constant of the
calibration record
Accept the cell constant of the last calibration record as
follows:
Selecting any cell
constant
You can set the cell constant manually in the range 0.250 ...
2.500 cm
-1
as follows:
1
Switch to the
Menu
mode.
2
Use
46
to select the main
Cal
menu.
3
Select the
Cell const.
menu item with
82
4
Confirm the
Cell const.
menu item.
5
Select the
Cal cell
menu item with
82
.
6
Confirm the
Cal cell
menu item with
g
.
The cell constant of the last calibration is accepted.
Cal cell
is
a
ctive (
✔
).
7
Terminate the
Menu
mode with
m
.
1
Switch to the
Menu
mode.
2
Use
46
to select the main
Cal
menu.
3
Select the
Cell const.
menu item with
82
4
Confirm the
Cell const.
menu item.
5
Select the
Man cell
menu item with
82
.