pH
79
Selecting the
calibration
procedure
Selecting the buffer
data record
For each calibration procedure, you can select between the
buffer datasets
TEC
,
DIN
and
Mrk
:
1
Switch to the
Menu
mode.
2
Use
46
to select the main
Cal
menu.
3
Select the
pH caltype
menu item with
82
.
4
Confirm the
pH caltype
menu item with
g
.
5
Use
82
to select the
AutoCal
or
ConCal
menu item.
6
Confirm the selected menu item with
g
.
The selected calibration procedure is active (
✔
).
7
Select the
pH Buffer
menu item with
82
.
8
Confirm the
pH Buffer
menu item with
g
.
9
Select the
Buffer TEC
,
Buffer DIN
or
Buffer Mrk 1...3
menu item with
82
.
10
Confirm the selected buffer data record with
g
. The
selected setting is active (
✔
).
11
Terminate the
Menu
mode with
m
.