![wtw inoLab 740 Operating Manual Download Page 129](http://html1.mh-extra.com/html/wtw/inolab-740/inolab-740_operating-manual_1000561129.webp)
Ion concentration
129
9
Start the measurement with
[Start]
.
10
Add the standard solution to the sample.
11
Start the measurement with
[Start]
.
12
Enter the
Standard volume
, and the
Ion type
for the
sample (#1) and standard solution (#2) with
82
and
[Next]
.
13
For the entry of a self-defining type of ion (
??
):
l
Select the
??
entry for the ion in the sample (#1)
with
82
.
l
Set the valency and molecular weight for the ion in
the sample with
[Next]
and
82
.
l
Select the
??
entry for the ion in the standard
solution (#2) with
[Next]
and
82
.
l
Set the valency and molecular weight for the ion in
the standard solution with
[Next]
and
82
.
14
Terminate the measurement with
[Start]
.
Summary of Contents for inoLab 740
Page 18: ...Safety 18...
Page 27: ...Commissioning 27...
Page 32: ...Commissioning 32...
Page 50: ...Basic principles of operation 50...
Page 92: ...pH 92...
Page 134: ...Ion concentration 134...
Page 148: ...Dissolved oxygen D O 148...
Page 174: ...BOD 174...
Page 232: ...Data output 232...
Page 236: ...Maintenance cleaning disposal 236...
Page 270: ...Lists 270...