What to do if...
240
No stable
measured value
Cause
Remedy
Measuring system:
– Incorrect calibration
procedure
– Select correct procedure
– Incorrect solution
temperature
(without temperature
probe)
– Set up correct
temperature
– Socket damp
– Dry socket
Buffer solutions
– Incorrect buffer solutions
– Change calibration
procedure
– Buffer solutions too old
– Use only once.
Note the shelf life
– Buffer solutions depleted
– Change solutions
Cause
Remedy
pH/ORP electrode:
– Diaphragm contaminated
– Clean diaphragm
– Membrane contaminated
– Clean membrane
Test sample:
– pH value not stable
– Measure with air
excluded if necessary
– Temperature not stable
– Adjust temperature if
necessary