Ion concentration
98
7.2.3
Reference measurement (
Ref Meas
)
A reference measurement enables you to assign the voltage
value 0 mV to the calibration curve of your electrode for a
specific ion concentration (reference concentration).
Prerequisite:
l
The measured value display must show the voltage.
l
A valid calibration must be available.
1
Switch to the
Display
mode.
2
Only if no valid calibration is available
(no sensor symbol on the measured value display):
Perform a calibration (see section 7.3).
3
Switch to the display of the voltage (
U
) with
m
.
4
Switch to the
Menu
mode.
5
Use
46
to select the main
Meas
menu.
6
Select the
ISE
menu item with
82
.
7
Confirm the
ISE
menu item with
g
.
8
Select the
ISE setup
menu item with
82
.
9
Confirm the
ISE setup
menu item with
g
.
10
Select the
Ref Mess
menu item with
82
.
11
Confirm the
Ref Meas
menu item with
g
.
The
Ref Mess
function is active (
✔
).
The
-
symbol flashes on the measured value
display.
12
Switch to the measured value display with
m
.
13
Open the input field for temperature and reference
concentration with
82
.
Summary of Contents for inoLab 740
Page 18: ...Safety 18...
Page 27: ...Commissioning 27...
Page 32: ...Commissioning 32...
Page 50: ...Basic principles of operation 50...
Page 92: ...pH 92...
Page 134: ...Ion concentration 134...
Page 148: ...Dissolved oxygen D O 148...
Page 174: ...BOD 174...
Page 232: ...Data output 232...
Page 236: ...Maintenance cleaning disposal 236...
Page 270: ...Lists 270...
Page 274: ...Lists 274...