XR3D-6 Maintenance & Troubleshooting
EM-XR3D-6-001-EN.D
3-25
©2011 VOTI Inc.
The DIOX window appears.
3. Press the XRAY button (
).
The lights on the top of the scanner will turn red. The DIOX indicators for Red/Green Lights
and X-Ray On/Off will also turn red.
It means that the source is enabled to generate X-Rays but it is not powered. Since the
source is not powered, there is no radiation emitted.
4. Start VOTI's Diagnostic Tool application, which is the software for the X-ray source, located
at
c:\MonoBlock XP\X3825 Spellman*.exe
.
5. Enter parameters if not already displayed.
6. Start XView using
C:\Program Files\DTB\X-view4.2\XView4.2.exe
.