The Conveyor Belts starts in a forward motion. The Motor ON/OFF indicator turns green and
the Motor Direction turns Green.
2. Click OFF (
).
The conveyor belt stops. The Motor ON/OFF turns RED.
3. Click REV (
).
The conveyor belt starts in a reverse motion. The Motor ON/OFF indicator turns RED and
the Motor Direction turns RED.
4. Click OFF (
).
The Conveyor Belt stops. The Motor ON/OFF and the Motor Direction indicators turn RED.
5. Turn the Operator Key Switch to the ON position.
The Keylock indictor turns Green. Green means the circuit is open and the key will not
enable the X-ray.
6. Turn the Operator Key Switch to the OFF position.
The Keylock Indicator turns RED. Red means the circuit is closed and the key is enabling
the X-ray generation (default state)
7. Place an item in the tunnel to block the Photo Sensor beam in the entrance of the tunnel.
The EC1 indicator turns RED. Red means there is an object blocking the beam and can start
the X-ray generator.
8. Removing the item turns the EC1 indicator GREEN.
Green means the circuit is open and the switch will not enable the X-ray generator.
9. Placing and removing an item in the tunnel to block the Photo Sensor beam in the exit of the
tunnel causes the EC2 Indicator to change color.
10. Open the Access Panels, (front or rear) causes the Door Interlock indicator to turn color.
Press them (or pull them) to see the colour change. Green means the circuit is open and the
switches do not enable the X-ray generator. Red means the circuit is closed and the
switches can enable the X-ray (default state).
11. Press the XRAY button (
).
The lights on the top of the scanner turn red. The DIOX indicators for Red/Green Lights and
X-Ray On/Off turn red. It means that the source is enabled to generate X-Rays but it is not
powered.
The source is not powered, therefore, there is no radiation emitted.
12. Click LOGOFF (
) to exit the DIOX application.
3.3 Control Box Access - Technician
The technician uses the Control Box to interact with VOTI’s BioSans software to control the
scanner functions and monitor all the peripheries of the XR3D. The technician can view status
and error information on the LCD display and use the keypad to change information.
Summary of Contents for XR3D-6
Page 2: ...XR3D 6 Maintenance Troubleshooting ii EM XR3D 6 001 EN D 2011 VOTI Inc...
Page 6: ......
Page 20: ...XR3D 6 Maintenance Troubleshooting 2 8 EM XR3D 6 001 EN D 2011 VOTI Inc...
Page 32: ...XR3D 6 Maintenance Troubleshooting 3 12 EM XR3D 6 001 EN D 2011 VOTI Inc...
Page 48: ...XR3D 6 Maintenance Troubleshooting 3 28 EM XR3D 6 001 EN D 2011 VOTI Inc...
Page 52: ...XR3D 6 Maintenance Troubleshooting 4 4 EM XR3D 6 001 EN D 2011 VOTI Inc...