3. Turn off the UPS by pressing the power button.
4. From the front access panel, unplug the X-ray source from the Power distribution unit.
5. Unscrew then unplug the Digital Signal Interface (JB5) connector at the X-ray Source.
6. Push the locking tab and unplug the 7-Pin Molex Connector at the X-ray Source.
7. From the back access panel, unbolt four (4) bolts securing the funnel bracket to the XR3D
and remove the bracket and, using work gloves, remove the lead shielding.
8. Remove four (4) bolts securing the X-ray Source base to the XR3D.
9. Using the X-ray Source handles, pull the X-ray Source out of the XR3D base.