61
Symptom
Cause
Remedial Action
[
] is displayed.
•
The balance was subjected to
shock or vibrations during
span adjustment.
→
Reperform the span
adjustment while taking care
to shield the balance from
shock or vibrations (15).
→
Contact our technical service
personnel or your local
dealer.
[
] is displayed.
•
Span adjustment or span test
was conducted with an object
loaded on the measurement
span.
→
Ensure that there is nothing
loaded on the measurement
pan, and reperform the span
adjustment or span test (15).
[
] is displayed.
•
The balance was subjected to
shock or vibrations during
span adjustment.
→
Reperform the span
adjustment while taking care
to shield the balance from
shock or vibrations (15).
→
Contact our technical service
personnel or your local
dealer.
[
] is displayed.
•
Malfunction of the built-in
calibration weight device.
→
Contact our technical service
personnel or your local
dealer.
[
] is displayed.
•
Malfunction of the built-in
clock.
→
Contact our technical service
personnel or your local
dealer.
[
] is displayed.
•
Counting mode: The samples'
unit weight was insufficient.
•
%-mode: Conducted with a
sample that is lighter than
the %-limit weight.
→
Conduct with a sample that is
heavier than the countable
unit weight.
→
Conduct with a sample that is
heavier than the limit weight
value (27).
[
] is displayed.
•
Addition mode
: Double addition was made.
: Negative addition was made.
: Zero was added.
→
Try again in the addition
mode (28).
[
] was displayed.
•
A weight error in excess of the
±
1.50 mg range was specified.
→
Check the weight to be used
(use a weight with an weight
error within the
±
1.50mg)
(34).
The number in parentheses ( ) denotes the reference page.