25
8.6 Counting Mode
Use this function after setting measurement mode to Counting Mode [
]. (See
“9. Functions of Function 1” on page 30.)
1. Sampling operation (Storage of unit weight
values)
(1) Press the Zero/Tare key to reset the indication to
zero.
Follow the same procedure when a container is used.
(1)
/
○
PCS
(2) Press the Set key. The display changes to flash
[
], which means that ten samples are
going to be loaded. Pressing the Zero/Tare key in
this condition allows you to change the number of
samples, as follows: 10
→
30
→
50
→
100
→
10.
*1
(3) Count the samples precisely, according to the
displayed number and load them on the container.
(2)
(3)
Samples loaded
○→0← PCS
↓
M
↓
(4) Press the Function key . Only the [M] mark
flashes for several seconds, and the saved number
of samples flashes.
*2
(5) Press the Set key. The unit weight values are
saved and you are returned to measurement mode.
(6) If additional samples are loaded, the total number of
samples loaded will be displayed.
(4)
M
↓
M
↓
PCS
M
(5)
PCS
M
↓
○ PCS
*1 The greater the number of samples, the more
averaged weight is saved, contributing to the
reduction of counting errors.
*2 If [
] appears in the operations of (3) and (4)
and, thereafter, the [
] mark lights up at the lower
left of the display, it indicates a potential for counting
errors due to lightweight samples. If that is the
case, executing the “Memory Update Method”
described in the following subsection will help
improve the counting accuracy.
(6)
Number of
loaded samples
displayed
PCS
Caution:
If the unit weights of samples are too light to conduct the counting operation, the balance will
display [
].