16
(4) End of span adjustment
When the setup of a weighing capacity point is over,
the balance displays [
], and returns you to
measurement mode.
CAL
○
g
Cautions:
1. While the adjustment is in progress, do not touch the measurement pan and protect the
balance from shock or movement. If the balance is moved, the processing may be halted at
the stage of the flashing [
].
2. To cancel the adjustment midway, press the Print key. The balance returns to
measurement mode after briefly displaying [
].
3. Avoid conducting the span adjustment immediately after the balance has been powered up.
Always wait for at least 30 minutes before conducting the adjustment.
7.3 Span Test with Built-In Weight
This function enables the user to conduct the span test automatically by simply pressing the
Cal key. Conduct this test after setting the function to Span Test with Built-In Weight
[
]. (See “9. Functions of Function 1” on page 30.) While the span test is in progress, a
[CAL] mark will flash.
* This function is not available on models without the built-in weight.
(1) Start of span test
Make sure that there is nothing loaded on the
measurement pan and that the windshield doors are
completely closed. Press the Cal key. The
balance displays [
] and initiates the span
test.
○
g
CAL
(2) Setup of zero-point
The display begins to a flash [
], and the
balance automatically checks the zero-point.
(3) Check of weighing capacity point
When the zero-point check is over, the balance now
displays [
]. Several seconds later, the
display flashes and the balance automatically checks
the weighing capacity point.
CAL
CAL
CAL
↓
(4) Display of span test result
When the weighing capacity point check is
completed, the balance displays [
] and then
the result (error) of the span test in a weight value.
CAL
g