36
10.4 Calibration of Built-In Weight
This function calibrates the weight incorporated in the balance with reference to another weight you
have
*1
. When conducting the calibration, set the error of the weight you are going to use.
* This function is not available on models without the built-in weight.
(1) Keep pressing the Function key while
holding down the Zero/Tare key. When
[
] appears, release the key. The
display changes to [
].
(2) Press the Function key twice to display
[
].
(1)
/
+
Key released
○
g
(3) Set the rightmost digit to [
] by pressing the
Zero/Tare key.
(2), (3)
(3) Press the Function key to display
[
]
*2
.
(4) Press the Function key while holding down
the Zero/Tare key, then release the key.
(4)
/
+
CAL
↓
(5)
Weight error
displayed
mg
+
↓
CAL
↓
Weight loaded
CAL
(6)
CAL
↓
Weight unloaded
CAL
(5) After the specified error of the weight is displayed
*3
,
[
] flashes, the display changes to
[
]. When [
] appears, load
the weight onto the measurement pan.
(6) [
] flashes and the display changes to
[
]. When [
] appears, unload
the weight.
(7) The display changes from flashing [
] to
[
], which also flashes for several
seconds. Then the balance displays [
]
and automatically transits to the following item
[
].
*1 For the calibration of the built-in weight, use a
weight of Class E
2
(or equivalent) or better under
OIML Standard.
*2 To cancel the calibration, press the Print key.
The following item [
] now appears.
*3 For details, refer to “10.3 Setup of Error of Weight
in Use” (page 34).
(7)
CAL
↓
CAL
↓