15
7. Calibration of Balance
7.1 Advice CAL
To ensure conduct accurate measurement under varying environmental conditions, it is essential to
calibrate the balance (span adjustment). “Advice CAL” denotes the function that indicates the
suggested timing of calibration (span adjustment) by causing the [CAL] mark to flash. If the [CAL]
mark starts flashing, execute calibration (span adjustment) of the balance as soon as possible.
(1) Display [
] by using “9. Functions of Function 1” (page 30) as the guide.
(2) Set the rightmost digit to [
]
*1
by pressing the Zero/Tare key.
(3) Press the Set key to go back to measurement mode.
*1 To deactivate the function, display [
].
7.2 Span Adjustment with Built-in Weight (Auto-Span Adjustment)
This function enables the user to practice the span adjustment automatically by simply pressing
the Cal key. This function is effective when the function is set to Span Adjustment with Built-in
Weight (Auto-Span Adjustment) [
]. (See “9. Functions of Function 1” on page 30.)
While the span adjustment is in progress, a [CAL] mark will flash.
* This function is not available on models without the built-in weight.
(1) Start of span adjustment
Make sure that there is nothing loaded on the
measurement pan, and that the windshield doors are
completely closed. Then, press the Cal key. The
balance displays [
] and, then, initiates the
span adjustment.
○
g
CAL
(2) Setup of zero-point
The display begins to flash [
], and the balance
automatically sets a zero-point.
CAL
(3) Setup of weighing capacity point
When the zero-point setup is over, the balance now
displays [
]. Several seconds later, the display
begins to flash and the balance automatically sets a
weighing capacity point.