39
11.4 Outputting Measurement Data in Compliance with ISO/GLP/GMP
When the statistical operation function of the printer is to be utilized, match the print characters
(English or Japanese) between the balance and the printer.
(1) Set the Print Control of the printer to the balance
side.
Continuous press
○
g
↓
(2) Refer to “9. Functions of Function 1” (page 30) and
verify that the item “Compliance with
ISO/GLP/GMP” is set to Enable ([
]),
and the item “Measurement Data Compliant with
ISO/GLP/GMP” is set to Comply ([
]).
(3) Output of header
Hold down the Print key. When [
]
appears, release the key. The header is printed.
After printing the header, the printer outputs the
measurement data.
Continuous press
○
g
↓
(4) Output of footer
Hold down the Print key. When [
]
appears, release the key. The footer is printed.
Caution:
The footer is output after the header is printed. However, if any changes are made in the item
“Compliance with ISO/GLP/GMP” or the item “Measurement Data Compliant with
ISO/GLP/GMP,” or if the balance is powered off, the print will resume from the header.