2. Memory Update Method (Improvement of Counting Accuracy)
If the unit weights of samples are so light that [
] is displayed, or a large number of articles
are going to be counted (more than five times the sampled number), executing the “Memory
Update Method” will help improve the counting accuracy.
PCS
M
PCS
M
PCS
M
PCS
M
(1) Perform the sampling operation. (See Steps (1)
through (4) in the preceding subsection.)
(2) Add a few times the current number of samples.
(No need to count.)
Press the Function key. The unit weight is
saved (updated), the number of samples flashes
again, and the Memory Update Method continues.
(3) Add more samples and repeat step (2). Save about
one half to one-fifth the quantity to be measured.
(4) Press the Set key. The balance saves the unit
weight values and restores measurement mode.
PCS
* Counting errors will decrease if you repeat the Memory Update Method as many times as
necessary, until the [
] mark at the lower left of the display disappears.
(Press the Set key to terminate sampling even during displaying of [
] mark.)
Caution:
If the unit weights of samples are too light to conduct the counting operation, the balance will
display [
].