12
5. Operation Check
5.1 Start
Σ
0
Press the On/Off key. All the indications start
flashing. Check the indications to ensure that no
indication is missing or unlit.
Only the [M] mark will flash for several seconds, and the
balance will change to measurement mode.
M
↓
○
g
5.2 Operation Check
Push the measurement pan lightly by hand and make
certain that the weight indication does vary and that it
returns to the original indication after you release your
hand. (If the zero-point is deviated, press the
Zero/Tare key to adjust the zero-point.)
○
g
↓
g
↓
○
g
* For the balance to perform stable measurement, start measurement after 30 minutes has
elapsed (warming-up) following the power-up. During this process, leave the right and left doors
open halfway to allow the interior of the windshield to be acclimatized to the external air
temperature. A temperature difference between the inside and the outside of the windshield
may give rise to convection when the doors are opened/closed, thereby preventing accurate
measurements from being conducted.