DOC: ATP-03-925110-XX
DESC:
PAGE:
REV
15 of 16
TEST & CALIBRATION
2
PREPARED BY: HOWIE PLITSCH REVISED BY:
CONTROL: INDUSTRIAL ENGINEERING
REV. DATE:
SHOWN
-
08 - 01 - 00
Sax Makthepharack
Fixture Leak
- Used to diagnose fixture leaks.
*
By jumping Carrier in to Carrier out, blocking off Pressure Sense and Jumping Split Vent to Vent and Capping off the Low and High vent of the Master Flow
Sensors, Leaks to the atmosphere can be determined.
*
Works similar to Module Leak, except it is used for fixture diagnostics.
*
Click on the arrow in the upper left hand corner to start.
*