DOC: ATP-03-925110-XX
DESC:
PAGE:
REV
16 of 16
TEST & CALIBRATION
2
PREPARED BY: HOWIE PLITSCH REVISED BY:
CONTROL: INDUSTRIAL ENGINEERING
REV. DATE:
SHOWN
-
08 - 01 - 00
Sax Makthepharack
Reference Flow Sensor Calibration
- Used to facilitate and check Master Flow Sensor calibration.
*
From the Pull down menu, select High or Low Sensor to check.
*
Enter the Password of "1951"
*
Push the Botton of "Read Low Flow" or "Read High Flow" as is appropriate.
*
Select a gas type.
*
Select a module (Typically a Type 1) to run.
*
Set the Master Flow Sensor "Serial Number"
*
Press "Run " to start.
*
To end, Set the Pull down to "exit" and press "run"