3800GC, Burn-In Test Procedure
2
PREPARED BY: Sax Makthepharack REVISED BY:
CONTROL: INDUSTRIAL ENGINEERING
REV. DATE:
SHOWN
-
07 - 30 - 99
F.Faili/M.Slaughter
1.10
DETECTOR EFC Flow rate check
1.10.1 Turn on all gas supplies
1.10.2 Plug all other detector inlets (top)
1.10.3 Press “Detector” key to get to the Detector Menu
1.10.4 For each detector:
*
Select Detector
*
Go to the “adjustments” page (2)
*
Set the individual channel flows to 0
*
Unplug the Detector Base (@ the column end), attach a flowmeter
*
Turn on the gas
*
For each channel of each Detector, verify that the flow measured is within the below listed specs:
Detector Type
Channel Gas
Min/Max
Flow (cc/min)
Range
FID
1
He 45/55
Ref. + 10% full scale
II
H 45/55
Air 450/550
2
3
TSD
1
He 45/55
H 5.4/6.6
Air
2
180/220
2
3
II
II
II
II
2
2
CONTINUE