3800GC, Burn-In Test Procedure
2
PREPARED BY: Sax Makthepharack REVISED BY:
CONTROL: INDUSTRIAL ENGINEERING
REV. DATE:
SHOWN
-
07 - 30 - 99
F.Faili/M.Slaughter
1.5.8
Verify Time & Date (see diagram).
1.5.9
Edit Heated Zones (type, temp. limit, location, coolant) for GSV’s, see labels & flow diagram for temp. limits (see diagram).
1.5.10 Edit Column Parameters (see diagram).
*NOTE: Use N
2
by default, except He must be used as carrier gas in certain applications such as Mass Spec units & TCD.
Carrier*
Length
Bore
Manual
N
2
or He
30
320
Type 1 EFC
N
2
or He
30
320
Type 5 EFC
N
2
or He
30
320
Type 3 EFC
N
2
or He
30
320
Type 4 EFC (GSV)
N
2
or He
30
320
1.5.11 Edit Valves (see labels & flow diagram) (see diagram)
1.5.12 Save & Exit
1.6
1077/1079 INJECTORS
1.6.1
Press Injector key to get to menu (see diagram)
1.6.2
Select Injector
1.6.3
Set temperature to 50
°
C
1.6.4
Page down & turn oven to ON
1.6.5
Page down & turn Split state to ON
1.6.6
Perform steps 1.6.1-1.6.5 for all 1077/1079 Injectors
1.7
1041/60/61 INJECTORS
1.7.1
Press Injector key to get to menu
1.7.2
Select Injector
1.7.3
Set temperature to 50
°
C
1.7.4
Set oven power to ON