DOC: ATP-03-925110-XX
DESC:
PAGE:
REV
11 of 16
TEST & CALIBRATION
2
PREPARED BY: HOWIE PLITSCH REVISED BY:
CONTROL: INDUSTRIAL ENGINEERING
REV. DATE:
SHOWN
-
08 - 01 - 00
Sax Makthepharack
Module Leak
- Tool used to pressurize individual modules and test manually for leaks.
*
Set Function to "Pressurize"
*
Select the "Type", "Module" and "Gas"
*
Press "Run". The Mensor Pressure should increase to about 90 psig.
*
In the Function, Select "Isolate", Press "Run". This will close off all of the valves to the EFC so a pressure degradation test can be observed (Mensor gage)
*
When test is over, Select "Vent" from the Function Menu and press "Run". This will depressureize the system.
*
When the system is fully vented, select "exit" from the function menu and press "Run". This will get you back to the main menu.