CP-3800 GC Operator’s Manual
21
Ethernet Setup
Ethernet address - can enter IP address, Netmask, and Default Gateway
when provided by your network administrator. If using Star WorkStation
allow the WorkStation BootP server to assign IP addresses.
ACTIVE
METHOD
EDIT
METHOD
CTRL
RUN
TIME
END
TIME
Method 1
Method 1
0.00
20.00
INSTRUMENT STATUS (FRONT), Page 1 of 3
Component
Setpoint Actual
1079 Oven (
°
C)
50
50
Column Flow (mL/min)
1.0
Column Pressure (psi)
15.0
15.0
Column Oven Temp (
°
C)
100
100
FID Oven (
°
C)
300
300
FID Output (mV)
8.25
Software Ver.
View Faults
The Software Ver. Softkey displays the current version of installed software.
The View Faults softkey displays the most recently detected fault conditions.
Instrument
MENU Key
The Instrument MENU has four sub categories: Remote / Local, Standby
Mode, 8400 Miscellaneous, and Column Mode. Each of these four items are
discussed below.
The Remote / Local screen allows you to select between remote (Star
Workstation) or local control. Disabling remote control prevents a remote
user from gaining control of the instrument, which may be an important
safeguard when conducting instrument maintenance. This screen also
displays the Ethernet Address, IP address, and Alias assigned by the remote
user. This information cannot be configured locally on the GC but has to be
assigned from the Star Workstation. Refer to the Communications section of
this manual for more detailed information on configuring the 3800 with a Star
Workstation.
The screen below shows a typical configuration where the 3800 is under
remote control. To disable remote control, move the cursor to the Enable
field and toggle to “Prevent” using the INCR or DECR key.