CP-3800 GC Operator’s Manual
17
Column
Parameters
Column Parameters defines the carrier gas and the capillary column
dimensions used with the specific sample introduction device (injector or
valve). The location (front, middle, or rear) refers to the Sample
Introduction device to which the column is connected. The carrier gas
and column dimension information is used for EFC calculations. If a
packed column is used, then “0” is set for length and ID. The
parameters and their ranges are shown below:
Default
Range
Location
Front
[Front, Middle, Rear]
Carrier Gas
He
[He, H
2
, N
2
]
Length (meters)
30.00
[0 – 250.00]
ID (
µ
m)
250
[0 - 999]
The following example is shown for a 3800 with two capillary columns installed.
ACTIVE
METHOD
EDIT
METHOD
CTRL
RUN
TIME
END
TIME
EDIT COLUMN PARAMETERS
Location
Carrier
Length ID
Front
He
30.00
250
Middle
He
30.00
320
Rear
He
0.00
0
Save and Exit