CP-3800 GC Operator’s Manual
13
Type 13
EFC
Type 13 EFC is a detector EFC module designed for the Electron Capture
Detector (ECD). It is a flow controlled system with one channel which
supplies make-up gas. The gas type can be set to Nitrogen (N
2
) or Argon
(Ar).
ACTIVE
METHOD
EDIT
METHOD
CTRL
RUN
TIME
END
TIME
EDIT FRONT EFC (Type 13)
Detector:
ECD
Channel 1 Gas Type:
He make-up
AutoCalibrate
Save and Exit
For this Parameter …
Enter…
Channel 1 Gas Type
He, N
2
or Ar.
Type 13 EFC has the following softkeys:
Press this softkey
If you want to…
AutoCalibrate Automatically
calibrate the EFC system. When it is finished, a
screen will display either a successful completion or detected
problem message. No other GC functions can be performed
during AutoCalibration.
Note that AutoCalibration should be done
on a semiannual basis, or whenever an EFC module is changed.
Save and Exit
Save the changed parameters and reinitialize the 3800.