The Standby Mode allows you to set a standby temperature for the column
oven. This standby temperature will be invoked after the timer counts to zero,
provided another run is not started before the timer has timed out. Once the
timer had timed out a method activation will restore the column temperature
to the method temperature. A timer value of zero disables the standby
temperature function.
8400 Miscellaneous allows you to: turn the autosampler buzzer On/Off,
change the plunger stroke warning limit, turn the vial detector On/Off, and
Find home at Start. On page two you can change the needle depth into the
injector and change the injection speed. These parameters are also
discussed in more detail in the 8400/8410 Operator Manual.
Column mode defaults to Standard. If this 3800 is part of a GC/MS system
Rapid MS mode can be selected from MS Workstation. Once selected, the
linear velocity will not be displayed and the column dimensions will be
dashed out in View Setup Column parameters.