To start a Single Method Automation, press the Select / Edit key in the
Automation section of the 3800 keyboard, then select Single Method as the
automation choice. Press the 3800 Start key to start automation. Note that if
the method specified in Single Method Automation is different from the
currently active 3800 method, the method specified in automation will be
activated upon pressing the Start key. Depending on the GC parameters of
the respective methods, this may result in a delay of several minutes before
the first analytical run starts.
Method Sequence, Addressable Automation
Method Sequence Automation allows running an automation sequence
where the method can be changed from run to run. The sequence of
methods is built by adding lines to the automation table. Unlike non-
addressable method sequences, the automation table may not be run
multiple times. The method sequence automation table can consist of up to
25 lines. Lines are added by pressing the New Line key on the 3800
keyboard. Note that for any given line, the final sample must be greater than
the initial sample. Prep-Ahead allows the autosampler to begin preparing the
next sample for injection while the current run is in progress. You will have to
experiment with the Prep-Ahead count down timer if you wish to have the
next sample ready to inject exactly when the GC becomes ready after the
current run ends.
ACTIVE
METHOD
EDIT
METHOD
CTRL
RUN
TIME
END
TIME
Method 1
Method 1
0.00
20.00
Page 1 of 2
METHOD SEQUENCE AUTOMATION
Method
Initial
Final
Inj / Vial
Inj Vol
Method 1
0
12
1
1.0
Method 2
13
24
1
1.0
Priority Sample
Priority Sample is only available while local Automation is running. You may
designate which vial location contains the priority and which Method (1-8)
you wish to use with it. Pressing the Run Now soft key will cause the
selected vial position to be sampled using the specified method