Below is the complete set of valve names and associated states that are
available for programming.
Valve Type
Abbreviation
Default State (-)
Energized State (+)
Front Split Valve
FSV
ON (split)
OFF (s/less)
Middle Split Valve
MSV
ON (split)
OFF (s/less)
Rear Split Valve
RSV
ON (split)
OFF (s/less)
Gas Sampling Valve
GSV
Fill
Inject
Liquid Sampling Valve
LSV
Fill
Inject
Sample S
OFF
ON
Internal Standard
IS
OFF
ON
Surrogate Standard
SS
OFF
ON
Series Bypass
SB
Series
Bypass
Backflush to Detector
BD Forward Backflush
Backflush to Vent
BV
Forward
Backflush
Column Selection
CS
Column 1
Column 2
Inj Backflush to Detector
IBD
Fill + Backflush
Inject
Inj Backflush to Vent
IBV
Fill + Backflush
Inject
Alternate Injection
AI
Column 1
Column 2
Simultaneous Injection
SI
Fill
Inject
Methanizer Bypass
MB
Series
Bypass
Sample Preconcentration Trap
SPT
SPT Desorb
SPT Trap
External Event A
EA
OFF
ON
External Event B
EB
OFF
ON
External Event C
EC
OFF
ON
External Event D
ED
OFF
ON
None
INJECTOR Key
The Injector section of the method can contain liquid injectors, valve ovens or
an SPT (Sample Preconcentration Trap). The Sample Preconcentration Trap
has a separate user's manual.
Liquid Injectors
Injector Description
CP-1177 Split/Splitless
Capillary
Dedicated Split/Splitless isothermal injector. Capillary
columns only.
1079 Universal Capillary
Split, splitless, and on-column injector. Temperature
programmable with cryogenic capability
Capillary columns only
1041 On-column
Isothermal, largebore and packed columns
1061 Flash Vaporization
Isothermal, largebore and packed columns