ELTS40 Elite Trendelenburg System 40° Set Up
1. Perform steps 1.2.4.5.6 from the ELTS20 Set Up,
excluding step 3. Straps do not need to be fastened.
2. Once the patient is correctly positioned, Shoulder
Supports can be added. To attach Shoulder Supports,
fit the L-Bracket provided by sliding the L-Bracket into
the receiver, located at the front of the support.
3. The L-Bracket should then be placed into the Clamp.
4. Gently slide the Shoulder Supports forward
underneath the patient.
5. If there is not a continuous rail, or if fixing prevents the
attachment, move the system back to a position that
best accommodates the Shoulder Supports and
Clamps.
6. Manoeuvre the pockets over the Shoulder Supports.
7. Secure the system by tightening the Clamps on the
surgical rail. Arm retainers can be added by pushing
the flat side of the board under the mattress
8. Before surgery a tilt test should be performed, to
ensure the patient is secure. Please note our system
also works in lateral tilt.
For further information and images please refer to the
ELTS20 & ELTS40 set-up guide included with your
products or view the set up video here
https://vimeo.com/180749129
EL913: Oasis Elite Lateral Park Bench Positioner
Ensure the gel side is facing up. Place the product with
the longer side under the patient’s body, ensuring the
shorter side is draped over the side of the OR table as per
image 3.
Please ensure straps are securely wrapped around the
side rails of the table
This product should be used in conjunction with the Ope-
rating Room Tables Positioning-specific accessories.
Please consult your table manufacturer for more informa-
tion.
The EL800: Oasis Elite Lateral Leg Positioner is also suita-
ble for this position. Ensure the patient is positioned sa-
fely and securely.
Oasis Elite Table Pads (EL030 and EL031)
For best results store product with Gel side facing
outwards.
Cleaning of your Trulife Pressurecare products.
• Products should be disinfected before and after use.
• Oasis, Oasis Elite, Azure and Oasis + products can be
easily cleaned using standard operating room cleaning
agents. Avoid soaking products in cleaning fluid.
• If product has Hook and Loop attachments keep it
closed when not in use to reduce the build up of dust.
• If the Oasis + product attracts fibers. It can be easily
cleaned under running water.
• Products should be dry before use.
• The Oasis + products have been tested for Autoclave
sterilisation. The tests reported no discernible change
from the products after 24 Autoclave Sterilisation
Cycles at 134 °C and 3 minutes.
• For the ELTS40 system please follow cleaning
instructions included with the Arm Retainer, the L-
Bracket, and the Clamps. (ELT22-ELT24) The Trulife
Shoulder Support (ELT21)can be cleaned using
standard operating room cleaning agents
For more detailed information on cleaning agents and
other testing, please see the technical guide in your
Trulife Pressurecare catalogue. (You can request this
from your local distributor)
Handling
Gently remove pads from tables or surgical equipment.
To transfer large table pads (not containing foam) you
may roll the pad and carry it cradled in arms. To transport
larger foam or gel products, fully support the pad with
two hands or cradle it in arms when possible. Carts are
recommended for transporting bulkier items.
Storage of your Trulife Pressurecare products
• Fold large table pads in sections of approximately
35cm.
• Avoid placing heavy loads or sharp objects on any of
the products.
• Please store all products flat.
• Store products at room temperature.
Disposal instructions
Any disposal practice must be in compliance with all
local and national laws and regulations.
All packaging materials used for your Pressurecare
product can be recycled.
In the unlikely event should any problems arise with
your Pressurecare product, please contact your local
distributor. If returning a damaged product, please
disinfect prior to sending.
If you have any questions regarding our product range
or their applications, or have any comments regarding
any Products, please contact your local distributor.