GX9 ASD Installation and Operation Manual
71
Virtual Linear Pump Setup
Toshiba International Corporation’s
Virtual Linear Pump
algorithm allows for direct and precise
control of pressure, flow rate, or level. This is achieved without the concerns, instabilities, or
complexities that are traditionally associated with pumping system control. This section provides useful
setup and operational information for the
Virtual Linear Pump
system.
The system is initially configured using the (Program
Virtual Linear Pump
Setup Wizard
. Once
the Wizard
is started it must be completed for normal
Virtual Linear Pump
operations to function.
However, the parameters addressed while using the Wizard or the
Virtual Linear Pump Settings
menu
selection are also accessible via their associated direct access numbers for specific adjustments when
required (see
If using direct access numbers or the
Virtual Linear Pump Settings
to configure the
Virtual Linear
Pump
algorithm, parameter
must first be set to
255:Virtual Linear Pump
to accept the new or
changed
Virtual Linear Pump
parameter settings. Upon completion of the parameter changes, set
parameter
1
or
2
to use the changed (or new) settings for normal
Virtual Linear Pump
operations (Zero may be selected at
to save the changes to be used later).
The setup procedure and the Wizard setup screens are shown below.
Figure 28. Input the Electrical Specifications of the Motor.
Figure 29. Input the Specifications of the Transducer.
etup Wizard
1. From the nameplate of the
motor, enter the FLA.
2. Select
Pressure
,
Flow
, or
Level
.
3. Select the command source;
EOI
,
VI/II
,
RR
,
Com. Opt
.
4. Set the
Low Frequency
Limit
. 15 Hz fits most appli-
cations.
5. Click
Next
to continue.
6. Set the unit of measure for
the transducer; pressure,
flow rate, or level (i.e., PSI,
GPM, Inches of Water
Column, Feet of Water
Column, or Cubic Feet per
Minute, °C, °F, or Custom).
7. Select the transducer output
signal type; current or
voltage and the range.
8. Set the maximum reading of
the transducer.
9. Set the minimum reading of
the transducer.
10.Click
Next
to continue.
Note:
The
Custom
selection allows for a user-
created unit of measure for the process
variable as measured by the
transducer.
Summary of Contents for GX9
Page 2: ......
Page 3: ...Document Number 66893 001 March 2014 GX9 ASD Installation Operation Manual ...
Page 4: ......
Page 12: ...iv GX9 ASD Installation and Operation Manual ...
Page 251: ...GX9 ASD Installation and Operation Manual 239 Figure 36 GX9 ASD 5 0 20 HP Enclosure ...
Page 252: ...240 GX9 ASD Installation and Operation Manual Figure 37 GX9 ASD 25 75 HP Enclosure ...
Page 253: ...GX9 ASD Installation and Operation Manual 241 Figure 38 GX9 ASD 100 200 HP Enclosure ...
Page 254: ...242 GX9 ASD Installation and Operation Manual Figure 39 GX9 ASD 250 400 HP Enclosure ...
Page 255: ...GX9 ASD Installation and Operation Manual 243 Figure 40 GX9 ASD 500 800 HP Enclosure ...
Page 256: ...244 GX9 ASD Installation and Operation Manual Figure 41 GX9 ASD 900 1200 HP Enclosure ...
Page 268: ......
Page 269: ......