Operation
Alarms Menu
Thermo Fisher Scientific
Model 15
i
Instruction Manual
3-61
INSTRUMENT CONFIGURATION:
>I/O EXPANSION BOARD YES
ZERO/SPAN VALVES YES
ZERO AIR SCRUBBER YES
WHEEL PURGE YES
DILUTION NO
RANGE AVG DIAGS ALARM
AUTO CALIBRATION NO
The Contact Information screen displays the customer service information.
●
In the Main Menu, choose Diagnostics >
Contact Information
.
CONTACT INFORMATION:
CALL CENTER: 508-520-0430
WEB: WWW.THERMO.COM
RANGE AVG DIAGS ALARM
The alarms menu displays a list of items that are monitored by the analyzer.
If the item being monitored goes outside the lower or upper limit, the
status of that item will go from “OK” to either “LOW” or “HIGH”,
respectively. If the alarm is not a level alarm, the status will go from “OK”
to “FAIL”. The number of alarms detected is displayed to indicate how
many alarms have occurred. If no alarms are detected, the number zero is
displayed.
To see the actual reading of an item and its minimum and maximum
limits, move the cursor to the item and press
.
Items displayed are determined by the options installed. The zero/span
check and auto calibration screens are visible only if the zero/span check or
auto calibration options are enabled. The motherboard status, interface
board status, and I/O Expansion board status (if installed) indicate that the
power supplies are working and connections are successful. There are no
setting screens for these alarms.
If any alarms are active, the alarm (bell) icon is displayed on the right side
of the status bar.
●
In the Main Menu, choose
Alarms
.
Contact Information
Alarms Menu