beyond either the minimum or maximum limit, an alarm is activated and
the alarm (bell) icon appears in the status bar on the Run screen and in the
Main Menu.
●
In the Main Menu, choose Alarms >
AGC Intensity
.
AGC INTENSITY:
ACTUAL 196646 Hz
>MIN 150000 Hz
MAX 300000 Hz
RANGE AVG DIAGS ALARM
The Minimum AGC Intensity alarm limit screen is used to change the
minimum sample flow alarm limit. The minimum and maximum AGC
Intensity screens function the same way.
●
In the Main Menu, choose Alarms > AGC Intensity >
Min
or
Max
.
AGC INTENSITY:
ACTUAL MIN: 150000 Hz
SET MIN TO: 200000 Hz ?
©ª
INC/DEC
SAVE VALUE
RANGE AVG DIAGS ALARM
The Motor Speed screen (read only) displays the current motor speed. A
reading other than 100.0% indicates a problem with either the motor or
the power supply.
●
In the Main Menu, choose Alarms >
Motor Speed
.
MOTOR SPEED:
ACTUAL 100.0 %
RANGE AVG DIAGS ALARM
The Zero and Span Check screens allow the user to view the status of the
most recent zero/span checks and set the maximum check offset. (The
Min and Max AGC
Intensity Limits
Motor Speed
Zero and Span Check