Operation
Instrument Controls Menu
Thermo Fisher Scientific
Model 15
i
Instruction Manual
3-45
The Select Range screen is used to select the hardware range for the selected
analog output channel. Possible ranges for the voltage outputs are: 0-100
mV, 0-1, 0-5, and 0-10 V. Possible ranges for the current outputs are: 0-20
mA and 4-20 mA.
●
In the Main Menu, choose Instrument Controls > I/O Configuration >
Analog Output Config > select Channel >
Select Range
.
SELECT OUTPUT RANGE:
SELECTED OUTPUT: V ALL
CURRENTLY: 0-10V
SET TO: 0-100mV ?
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
The Minimum and Maximum Value screens are used to edit the zero (0)
and full-scale (100) values, respectively, in percentages for the selected
analog output channel. See
Table 3–6
for a list of choices. The minimum
and maximum output value screens function the same way. The following
example shows the set minimum value screen.
●
In the Main Menu, choose Instrument Controls > IO Configuration >
Analog Output Config > select Channel >
Set Minimum
or
Maximum
Value
.
MINIMUM OUTPUT PERCENT:
SELECTED OUTPUT: V ALL
CURRENTLY: N/A %
SET TO: 0000.
5
% ?
©ª
CHANGE VALUE
SAVE
RANGE AVG DIAGS ALARM
Select Range
Minimum and Maximum
Value